Description
The Infant Incubator YR02186 is designed to provide a controlled environment that supports the health and development of newborns by maintaining optimal air and skin temperature. With servo-controlled temperature systems that integrate advanced computer technology, this incubator ensures precise thermal stability. Key features include various alarms, free step mattress tilting adjustment, an easy-to-clean removable humidity trough, and a humidity display and control function. Additionally, it offers an optional phototherapy unit, a guardrail for added safety, and comes equipped with a transfusion shelf and tray.
Market Price
Understanding the investment in a high-quality infant incubator is crucial for medical facilities. With the Infant Incubator YR02186, you can expect a competitive price range that starts from approximately USD 4600 to USD 5300. This reflects the advanced features and reliability that the incubator offers, making it a worthwhile investment for any healthcare environment.
Frequently Asked Questions
- What kind of alarms does the Infant Incubator YR02186 have? The incubator includes various self-check alarms for temperature and humidity to ensure the infant’s environment remains safe and stable.
- Is the phototherapy unit included? The phototherapy unit is available as an optional accessory for added neonatal care.
- How do I adjust the mattress? The incubator features a free step mattress tilting adjustment, allowing for tailored positioning of the infant.
Advantages and Disadvantages
The Infant Incubator YR02186 offers significant advantages, such as precise temperature control, ease of cleaning, and flexibility with optional accessories. A potential disadvantage might be the initial cost; however, the myriad of features and reliable performance justify the investment.
Field Use
In practical settings, the Infant Incubator YR02186 is indispensable for NICUs, providing a safe, controlled environment for neonates requiring special care. Its ease of use and adjustable features allow healthcare professionals to focus on providing the best care possible.
Recommendations
To ensure longevity and optimal performance, regularly clean the removable humidity trough and conduct routine checks on the alarm systems. Additionally, consider utilizing the optional accessories to tailor the incubator to specific needs, thereby maximizing its functionality.
Features
- Air and skin temperature servo-controlled by computer
- Various self-check alarms for safety
- Free step mattress adjustment
- Removable humidity trough for easy maintenance
- Includes transfusion shelf, tray, and guardrail
- Humidity display and control
- Optional phototherapy unit
Technical Specifications
Model | YR02186 |
Electrical requirement | ~220V 59Hz or ~230V 60Hz |
Power consumption | ≤450VA |
Air mode temperature range | 25.0°С~37.0°C |
Skin mode temperature range | 25.0°C~37.0°C |
Air over temperature range | <38.0°C |
Air mode high temperature range | 37.1°C~39.0°C |
Skin mode high temperature range | 37.1°C~38.0°C |
High temperature mode alarm | < 40.0°C |
Air mode alarm | ±3.0°C |
Skin mode alarm | ±1.0°C |
Skin sensor accuracy | ≤±3.0°C |
Air control accuracy | ≤0.5°C |
Mattress temperature uniformity | ≤0.8°C |
Noise | ≤55dB (A) |